Transforming retinal photocoagulation — effectively, efficiently
Transforming retinal photocoagulation — effectively, efficiently
Appointment.
With Integre Pro ScanTM from Ellex, you can transform the efficacy and efficiency of retinal photocoagulation. At the touch of a button, the system can deploy a comprehensive choice of patterns that are spaced, shaped and positioned consistently and precisely for the accurate and efficient treatment of retinal disease.
Description
Uniform energy distribution
Integre Pro Scan’s proprietary dual-mode laser cavity delivers uniform energy distribution across the full spot diameter, eliminating hotspots and achieving optimal, homogenous burns.
It’s a key feature that means you can achieve consistent, predictable treatment outcomes across a broad range of pathologies.
Integre Pro ScanTM from Ellex combines multi-color photocoagulation with a precise computer-controlled pattern scanning laser in an ergonomically advanced all-in-one laser/slit lamp design.
Whether positioning focal treatment in the macular area, or performing PRP in the periphery, Integre Pro ScanTM provides a comprehensive pattern and wavelength choice to cover all retinal pathologies.
Key Features
True Spot Optics
Integre Pro Scan’s enhanced
True SpotTM optical system
gives you a clearer view of the fundus. It’s a system that delivers excellent visualization and illumination, alongside enhanced depth perception and a wider peripheral view.
When it’s combined with an optimal red-reflex, it’s also a solution that enables you to titrate power levels precisely to deliver highly accurate treatment.
Integre Pro Scan’s user-friendly touchscreen interface means you can adjust all treatment parameters, including laser power, pulse duration, pattern selection and pattern size quickly and easily.
Integre Pro |
|
Laser Source |
Solid-state laser diode and cavity |
Wavelength |
1. Yellow-red configuration: 561 nm and 670 nm |
Power at the Cornea |
Red: 50 – 1000 mW, |
Exposure Time |
Pattern: 10, 20, 30, Single Spot: 0.01 to 8.0 seconds |
Pattern |
Linear (2 to 7 spots) |
Spot Size |
Pattern: 100 to 500 μm, continuously variable |
Spot Spacing |
Pattern: 0 (touching), |
Pulse Duration |
Pattern: 10, 20, 30 ms |
Repeat Mode |
Pattern: N/A, Single Spot: 50 – 1000 ms |
Aiming Beam |
Red 635 nm, adjustable intensity |
Magnification |
6x, 10x, 16x, 25x, 40x |
Electrical Requirements |
100-240 VAC, 50/60 Hz, 800 VA |
Cooling |
Air cooled |
Weight |
35kg, 77 lbs. (laser only) |
Dimensions (HxWxD) |
62 x 76 x 47 cm, 24 x 30 x 19 inches (laser only) |